产品简介:
902型小动物立体定向仪最初由David Kopf Instruments于1963年设计,是一种用途广泛、易于使用的仪器,用于电极、微管、套管和其他装置的立体定向,使小动物能够正确地对齐。
产品特点:
质量、多功能性和执行能力得到保证;
KOPF仪器;
结合学生水平的操作简单性;
具有研究质量和准确性;
允许和可靠的立体定位;
目前使用的几个图集是基于Kopf模型900的可靠参数;
模块化的配件允许扩大仪器的功能;
适用于各种小动物。
产品型号:
The instrument consists of a rugged “U” frame to which is affixed a Model 960 three-dimensional Electrode Manipulator. Model 1770 Standard Electrode Holder with corner clamp is included.
Manipulator X,Y, Z adjustment – Metric scale 80 mm travel, calibrated 0.1 mm vernier scale.
Angle adjustment – Fully universal joint calibrated on two planes for access from any angle. Vertical alignment pin can be removed for angled settings from 0° – 90° either side of vertical, 2° increments.
Rotation adjustment – Manipulator swivel base can be rotated up to 360°. Manipulator X/Z axis can be repositioned at 90° increments.
The stereotaxic frame is mounted to a 17″L x 10″W base plate and is elevated to bring ear bar height to 53 mm. Ear bar slots on the “U” frame have 0.1 mm verniers for ease in centering the animal. Ear bar zero to posterior end of base plate is 8.763″.
Model 920 Rat Adaptor features include a nose and tooth bar assembly offering various stereotaxic adjustments.
Dorsal/Ventral adjustment dial – Calibrated in 100 micron increments, 1 mm per revolution.
Dorsal/Ventral adjustment – 30 mm
Dorsal – 10 mm above ear bar zero
Ventral – 20 mm below ear bar zero
Adaptor Plate A/P adjustment – 44 mm
Ear Bars – Model 957 18° Tip Rat Ear Bars
(6.3 mm square with 35 mm of calibration).
Our instruments were designed for versatility. Any standard accessory, that is not required, can be omitted and the appropriate accessories can be added for your specific research.
Model 902 Standard Accessories
(supplied with Model 902)
Model 900 Stereotaxic “U” frame Assembly
Model 900-C base Plate
Model 920 Rat Adaptor including one set of
Model 957 18° Tip Rat Ear Bars
2-Model 960 Electrode Manipulator (0.1 mm res.) including
Model 1770 Electrode Holder with corner clamp
Our instruments were designed for versatility. Any standard accessory that is not required can be omitted and the appropriate accessories can be added for your specific research.